Patient-circumference-adapted dose regulation in body computed tomography. A practical and flexible formula
U Nyman1, T L Ahl, M Kristiansson
1Department of Diagnostic Radiology, Lasarettet Trelleborg, Sweden. ulf.nyman@skane.se
Purpose:
To illustrate that the attenuation formula based on monochromatic radiation in homogeneous objects may be used for dose regulation in body computed tomography (CT) based on patient circumference and using a simple cloth measuring tape.
Material And Methods:
Based on the attenuation formula for monochromatic radiation the following Microsoft Excel equation was derived: mAs(x) = mAs(n)*EXP((0.693/ HVT)*(O(x)-O(n))/PI()), where mAs(x) (milliampere second) in a patient with circumference O(x) is calculated based on the nominal mAs(n) set for a reference patient with the circumference O(n) with regard to indication, scan protocol, and available CT scanner. The HVT = half-value thickness (object thickness change in cm affecting mAs setting by a factor of 2) resulting in the least mAs difference compared with published studies investigating the mAs needed for constant image noise in abdominal CT phantoms at 80-140 kVp was evaluated. Clinically recommended HVT values were applied to 20 patients undergoing abdominal CT using 130 effective mAs and 94 cm circumference as nominal settings, and an HVT of 9 cm.
Results:
The object-sized dependent mAs for constant image noise at 80-140 kVp in 10-47 cm diameter abdominal phantoms (31-148 cm in circumference) differed, with few exceptions, by no more than 10% from those obtained with our formula using an HVT of 3.2-3.8 cm. An HVT of 9 cm in the patient study resulted in the same image noise-patient circumference relation as a phantom study using a "clinically adapted mAs" resulting in an acceptable noise according to diagnostic requirements. Clinical experiences recommend an HVT of about 8 cm for abdominal CT and 12 cm in thoracic CT. Changing the kVp from 120 to 80, 100, or 140 requires a mAs change roughly by factors of 4, 2, and 0.6, respectively, for constant image noise.
Conclusion:
Until fully automatic automatic exposure control systems have been introduced, applying the formula in a computer program provides the radiologist with an easy, quick, flexible, and practical instrument for reasonably good patient-sized adjusted exposure levels in clinical practice.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dosage Regimen Designs: Nomograms and Tabulations
Dosage Regimens: Designs and Approaches
Dosage Regimens: Partial Pharmacokinetic Parameters
Drug Dosing: Infants and Children
Dosage Regimen: Individualization