Variability among five over-the-counter blood glucose monitors
Mary M Kimberly1, Hubert W Vesper, Samuel P Caudill
1Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Mailstop F25, 4770 Buford Hwy NE, Atlanta, GA 30341-3724, USA. mkimberly@cdc.gov
Clinica Chimica Acta; International Journal of Clinical Chemistry
|September 7, 2005
Summary
Blood glucose monitor results vary significantly, impacting diabetes self-management. Standardization is crucial for accurate blood glucose concentration readings and improved patient care.
Area of Science:
- Clinical Chemistry
- Medical Devices
- Diabetes Management
Background:
- The American Diabetes Association recommends self-monitoring of blood glucose concentration for diabetes management.
- Variability among common blood glucose monitors necessitates an assessment for standardization.
Purpose of the Study:
- To evaluate the variability among five common blood glucose monitors.
- To determine the need for standardization in blood glucose monitoring devices.
Main Methods:
- Minimized preanalytical and user variability by using a single trained operator and standardized procedures.
- Tested each monitor twice per participant using both aged and fresh test strips.
- Analyzed monitor variability using ANOVA across different concentration ranges and strip lots.
Main Results:
- No statistically significant differences in total or within-strip lot coefficients of variation (CVs) were found among monitors.
- Statistically significant differences in among-strip lot CVs were observed, ranging from 0% to 7.5%.
- The degree of significance for among-strip lot CVs increased with higher blood glucose concentration ranges.
- Over half of paired monitor comparisons showed statistically significant average percent differences, ranging from 5.7% to 32.0%.
Conclusions:
- Blood glucose monitor results exhibit significant variability, leading to poor agreement among devices.
- Standardization of blood glucose monitoring devices is essential to reduce variability.
- Standardization will ultimately improve the quality of patient care in diabetes management.
Related Concept Videos
Variability: Analysis
Measures of variability are statistical metrics that reveal the dispersion pattern within a dataset. They are pivotal in biostatistics, providing insights into the heterogeneity within health and biological data. Variability signifies the degree to which data points diverge from one another, helping researchers understand the potential range of values and associated uncertainty within the data.
The range is a simple measure of variability, indicating the difference between the highest and...
The range is a simple measure of variability, indicating the difference between the highest and...
Hyperglycemia
Hyperglycemia is an abnormally high blood glucose level. It is diagnosed by fasting glucose ≥126 mg/dL, 2-hour oral glucose tolerance test (or OGTT) ≥200 mg/dL, random glucose ≥200 mg/dL with symptoms, or HbA1c ≥6.5%. However, HbA1c results may be unreliable in certain conditions, such as anemia or hemoglobinopathies, and the diagnosis should be confirmed unless classic symptoms are present. Postprandial hyperglycemia is typically considered significant when glucose levels exceed 180 mg/dL two...