Use of self-inflating bags for neonatal resuscitation
Sam Oddie1, Jonathan Wyllie, Andrew Scally
1Consultant Neonatologist, Bradford Royal Infirmary, Duckworth Lane, Bradford BD9 6RJ, UK. Sam.Oddie@bradfordhospitals.nhs.uk
Background:
Lung inflation is the most important, and most difficult step in newborn resuscitation. A wide variety of devices are used to achieve lung inflation, but there are relatively few data to guide clinicians in their choice of device.
Methods:
We tested the ability of instructors and trained candidates on a newborn life support course to deliver initial inflation breaths to a test lung, using a pressure limited blow-off valve, a 240-ml self-inflating bag and a 500-ml self-inflating bag in sequence.
Results:
Use of a 240-ml self-inflating bag was associated with shorter initial inflations of 1.8 s mean (95% CI 1.60-1.99 s), compared with 2.42 s (2.24-2.61 s), 2.40 s (2.08-2.71 s) for 500-ml self-inflating bags and "Tom Thumb" T piece, respectively. Delivery of breaths within a target pressure range of 30+/-5 cm H2O was significantly better using a T piece than either self-inflating bag (proportion within target range 0.05 (95% CI 0-0.11), 0.17 (95% CI 0.12-0.23), 0.89 (95% CI 0.83-0.94) for 240-ml and 500-ml self-inflating bags and "Tom Thumb" T piece, respectively. Excessive pressure delivery with both sizes of self-inflating bag was frequent.
Conclusions:
These data do not support use of 240-ml or 500-ml self-inflating bags for resuscitation of newborn term infants. A variable pressure T piece blow-off system may be the easiest device to use for newborn resuscitation and the most reliable at delivering desired pressures for set times.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)