Observation and absolute frequency measurements of the 1S0-3P0 optical clock transition in neutral ytterbium
C W Hoyt1, Z W Barber, C W Oates
1National Institute of Standards and Technology, 325 Broadway, Boulder, Colorado 80305, USA. hoycha@boulder.nist.gov
Abstract:
We report the direct excitation of the highly forbidden (6s2) 1S0 <--> (6s6p) 3P0 optical transition in two odd isotopes of neutral ytterbium. As the excitation laser frequency is scanned, absorption is detected by monitoring the depletion from an atomic cloud at approximately 70 microK in a magneto-optical trap. The measured frequency in 171Yb (F=1/2) is 518,295,836,591.6 +/- 4.4 kHz. The measured frequency in 173Yb (F=5/2) is 518,294,576,847.6 +/- 4.4 kHz. Measurements are made with a femtosecond-laser frequency comb calibrated by the National Institute of Standards and Technology cesium fountain clock and represent nearly a 10(6)-fold reduction in uncertainty. The natural linewidth of these J=0 to J=0 transitions is calculated to be approximately 10 mHz, making them well suited to support a new generation of optical atomic clocks based on confinement in an optical lattice.
Related Concept Videos
Atomic Nuclei: Larmor Precession Frequency
Atomic Spectroscopy: Absorption, Emission, and Fluorescence
Atomic Emission Spectroscopy: Instrumentation
UV–Vis Spectroscopy: Molecular Electronic Transitions
Atomic Emission Spectroscopy: Lab
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...