Updated: Aug 15, 2026

Magnetic Resonance-Guided Stereotaxy for Infusions to the Pig Brain
Published on: March 31, 2023
High-resolution real-time spiral MRI for guiding vascular interventions in a rabbit model at 1.5 T
Masahiro Terashima1, MinSu Hyon, Erasmo de la Pena-Almaguer
1Division of Cardiovascular Medicine, Stanford University School of Medicine, Stanford, California 94305-5233, USA.
Purpose:
To study the feasibility of a combined high spatial and temporal resolution real-time spiral MRI sequence for guiding coronary-sized vascular interventions.
Materials And Methods:
Eight New Zealand White rabbits (four normal and four with a surgically-created stenosis in the abdominal aorta) were studied. A real-time interactive spiral MRI sequence combining 1.1 x 1.1 mm(2) in-plane resolution and 189-msec total image acquisition time was used to image all phases of an interventional procedure (i.e., guidewire placement, balloon angioplasty, and stenting) in the rabbit aorta using coronary-sized devices on a 1.5 T MRI system.
Results:
Real-time spiral MRI identified all rabbit aortic stenoses and provided high-temporal-resolution visualization of guide-wires crossing the stenoses in all animals. Angioplasty balloon dilatation and deployment of coronary-sized copper stents in the rabbit aorta were also successfully imaged by real-time spiral MRI.
Conclusion:
Combining high spatial and temporal resolution with spiral MRI allows real-time MR-guided vascular intervention using coronary-sized devices in a rabbit model. This is a promising approach for guiding coronary interventions.