A comparison of a rapid test for influenza with laboratory-based diagnosis in a paediatric population

Robert Alexander1, Aeron C Hurt, David Lamb

  • 1Royal Children's Hospital, Parkville, Victoria.

Insights

The Directigen Flu A+B test shows 100% specificity for detecting influenza A and B but has 80.8% sensitivity. This rapid diagnostic offers potential patient management benefits, particularly with positive results.

Area of Science:

  • Clinical diagnostics
  • Virology
  • Pediatric infectious diseases

Background:

  • Accurate influenza A and B detection is crucial in hospitals for diagnosis, differential diagnosis, and guiding antiviral treatment.
  • Rapid diagnostic tests are valuable tools for timely patient management in pediatric settings.

Purpose of the Study:

  • To evaluate the performance of the Directigen Flu A+B rapid diagnostic test.
  • To compare its accuracy against established laboratory methods for influenza A and B detection in pediatric respiratory specimens.

Main Methods:

  • A total of 193 respiratory specimens were analyzed.
  • The Directigen Flu A+B test was compared with direct immunofluorescence, viral culture, and reverse transcriptase-polymerase chain reaction (RT-PCR).

Main Results:

  • The Directigen Flu A+B test demonstrated a sensitivity of 80.8% and a specificity of 100%.
  • Results align with previous pediatric studies indicating lower sensitivity compared to traditional laboratory tests.

Conclusions:

  • The Directigen Flu A+B test is highly specific for influenza A and B detection.
  • Despite lower sensitivity, the test has a potential role in patient management, especially when a positive result is obtained in pediatric settings.