Related Experiment Video
Updated: Aug 9, 2026

Use of Electromagnetic Navigational Transthoracic Needle Aspiration (E-TTNA) for Sampling of Lung Nodules
Published on: May 23, 2015
Computed-tomography-guided punctures using a new guidance device
A Magnusson1, E Radecka, M Lönnemark
1Department of Diagnostic Radiology, Uppsala University Hospital, Uppsala, Sweden. anders.magnusson@akademiska.se
Purpose:
To evaluate a new adjunctive guidance device, a puncture guide, constructed to simplify computed tomography (CT)-guided punctures and to make the procedure more accurate and safe.
Material And Methods:
17 patients referred for CT-guided punctures were included in the study. There were 10 thoracic and 7 abdominal or pelvic lesions with a mean maximum diameter of 29 +/- 18 mm. All punctures were performed using a laser guide combined with the new device. The needle guide created a streak artefact in the image, indicating the needle path.
Results:
The puncture was successful at the first attempt in 15 of the 17 patients. The artefact was visible in all patients, and in the majority there was a distinct artefact reaching from the entry point to the lesion. The deviation between the angle of the streak artefact and the final angle of the needle was 1.1 degrees.
Conclusion:
The benefits of the puncture guide were the artefact pointing at the target, the needle support, and accuracy when performing CT-guided punctures.