Related Experiment Video
Updated: Aug 15, 2026

Application and Methodology of the Non-destructive 19F Time-domain NMR Technique to Measure the Content in Fluorine-containing Drug Products
Published on: August 22, 2017
[Determination of ciprofloxacin in human serum by high-performance liquid chromatography]
1Service de biochimie et de pharmacologie clinique, Hôpital d'instruction des armées Percy, Clamart. sebastien.bardot@wanadoo.fr
Abstract:
A high-performance liquid chromatography method was developed for the rapid determination of ciprofloxacin in human serum. After protein precipitation, chromatographic separation of ciprofloxacin in plasma was achieved with an octyl column and 20 mmol/L phosphate buffer, methanol and acetonitrile 76/12/12 (v/v/v), pH 4, as mobile phase. Quantitative determination was monitored on a fluorescence detector using an excitation and emission wavelength of 278 and 450 nm, respectively. Enrofloxacin was used as internal standard. Calibration curve is linear over the range 0,12 to 20 microg/mL (r=0,9992). Limit of detection and limit of quantification are 0,04 and 0,12 microg/mL, respectively. Within-day coefficient of variation (2,0 to 3,1 %) and day-to-day coefficient of variation (4,3 to 5,5 %) at three different concentrations are sufficient. Relative recovery ranged from 91,1 to 94,0 % at three different concentrations. Each analysis requires no longer than 7 min.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors