Related Experiment Video
Updated: Aug 15, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Monte Carlo dosimetric study of the BEBIG Co-60 HDR source
F Ballester1, D Granero, J Pérez-Calatayud
1Department of Atomic, Molecular and Nuclear Physics and IFIC, University of Valencia-CSIC, Dr Moliner 50, E46100 Burjassot, Spain. Facundo.Ballester@uv.es
Abstract:
Although not as widespread as Ir-192, Co-60 is also available on afterloading equipment devoted to high dose rate brachytherapy, mainly addressed to the treatment of gynaecological lesions. The purpose of this study is to obtain the dosimetric parameters of the Co-60 source used by the BEBIG MultiSource remote afterloader (BEBIG GmbH, Germany) for which there are no dosimetric data available in the literature. The Monte Carlo code GEANT4 has been used to obtain the TG43 parameters and the 2D dose rate table in Cartesian coordinates of the BEBIG Co-60 HDR source. The dose rate constant, radial dose function and anisotropy function have been calculated and are presented in a tabular form as well as a detailed 2D dose rate table in Cartesian coordinates. These dosimetric datasets can be used as input data and to validate the treatment planning system calculations.