Rapid T1 measurement via decay-recovery decomposition: applications in fringe field and distributed relaxation
Eric E Sigmund1, Nicolas Caudal, Yi-Qiao Song
1Schlumberger-Doll Research, 36 Old Quarry Road, Ridgefield, CT 06877, USA.
Abstract:
Spin-lattice relaxation time (T(1)) measurements are often time-consuming due to the need to measure the full equilibrium magnetization with a long wait time. However, any magnetization recovery can be decomposed into pure recovery and pure decay components, the latter of which lends itself to a much simpler and faster extraction of T(1). We demonstrate several pulse sequences that accomplish this decomposition experimentally and illustrate its applications in a steady magnetic field gradient, and in materials possessing a broad distribution of T(1).
Related Concept Videos
Atomic Nuclei: Types of Nuclear Relaxation
In spin–lattice or longitudinal relaxation, the excited spins exchange energy with the surrounding lattice as they return to the lower energy level. Among several mechanisms that contribute to spin–lattice relaxation, magnetic dipolar interactions are significant. Here, the excited nucleus transfers energy to a nearby...
Protein Dynamics in Living Cells
Fluorescent recovery after photobleaching (FRAP) is a fluorescent-protein-based detection technique used to quantify protein movement rates within the cell. This method exposes a small portion of the cell to an intense laser beam. The laser beam causes permanent photobleaching of the fluorophore-tagged proteins in the exposed region. As the bleached...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Fast Reactions
NMR Spectrometers: Resolution and Error Correction