Integrated Compensatory Responses in a Human Model of Hemorrhage
Published on: November 20, 2016
Physiological evaluation of the U.S. Army one-handed tourniquet
Joseph C Wenke1, Thomas J Walters, Dominique J Greydanus
1U.S. Army Institute of Surgical Research, Fort Sam Houston, TX 78234-6315, USA.
Objective:
To provide a physiological assessment of the U.S. Army one-handed tourniquet (OHT).
Methods:
An OHT was self-applied by 26 subjects, to maximal tolerable tightness, to the proximal arm or thigh under different conditions and positions, and the presence of blood flow was assessed using Doppler ultrasonography or occlusion plethysmography.
Results:
Doppler sound was eliminated at the radial artery for all subjects with OHT application but was not stopped at the popliteal or dorsalis pedis artery for any subjects. The OHT reduced forearm blood flow by 79% but decreased leg blood flow by only approximately 50%, regardless of condition and position of application to the thigh.
Conclusions:
The OHT appears to effectively minimize blood flow in the arm but not in the lower extremities, and clinical assessment of blood flow disappearance by Doppler ultrasonography may underestimate the magnitude of actual blood flow to the limb.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: