Background:
Intracoronary beta-radiation therapy reduces in-stent restenosis (ISR). We aimed to determine the safety and feasibility of intracoronary radiation therapy (IRT) utilizing tungsten (188W), a beta emitter.
Methods:
A total of 30 patients with angiographic evidence of ISR in a previously treated native coronary artery underwent percutaneous coronary intervention (PCI; balloon angioplasty, ablation by atherectomy, or laser angioplasty). After the intervention, a noncentered delivery catheter with a side guide 0.014-in. wire carrying a tungsten (188W) coil, with an active length of 33 mm, was inserted. Patients were randomized to a radiation dose of 18, 22, or 25 Gy at 2 mm from the center of the source. Aspirin and Plavix, at 300 mg loading dose, were administered prior to intervention. Plavix 75 mg/day was prescribed for 6 months after the procedure.
Results:
At 6 months follow-up, the overall binary angiographic restenosis rate was 18.8%. Target vessel revascularization (TVR) was 23% and target lesion revascularization related major adverse cardiac events (TLR-MACE) was 13.3%, without any intergroup differences. A comparison with the original Washington Radiation for In-stent restenosis Trial (WRIST) radiation cohort utilizing an 192Iridium source (prescription dose 15 Gy at 2 mm from the source) showed similar TVR and TLR-MACE rates of 30% and 18%, respectively. The TVR and TLR-MACE rates in the WRIST placebo cohort were 70% and 66%, respectively.
Conclusions:
Vascular brachytherapy with tungsten (188W) is feasible and safe. The 6-month clinical outcomes are similar to the original WRIST radiation group.