Evaluation of a computed radiography system for megavoltage photon beam dosimetry
1University of Southern California and Childrens Hospital Los Angeles, Radiation Oncology Program, 4650 Sunset Boulevard, Los Angeles, California 90027, USA. aolch@chla.usc.edu
Abstract:
Computed radiography (CR) systems have been gaining adoption as digital replacements for film for diagnostic and therapy imaging. As a result, film processors are being removed from service, leaving a void for the medical physicists who use film and processors for two-dimensional mega-voltage beam dosimetry. This is the first report to evaluate the ability of a commercial CR reader and storage phosphor plate system to accurately quantitate absolute dose and dose distributions from a 6 MV photon beam. There are potential advantages and disadvantages of current CR systems compared to film systems. CR systems inherently produce a linear dose-response over several logs of dose. However, the barium in the storage phosphor has a higher atomic number than the silver in film, resulting in significant energy sensitivity. The purpose of this work is to fully characterize the impact of these and other features of this CR system relevant to dosimetry. The tests performed and reported on in this study include uniformity of readout across a uniform field, geometrical accuracy, intra- and interday reproducibility, signal decay with time and with light exposure, dose-to-signal calibration, high dose effects, obliquity effects, perpendicular and parallel calibration results, field size and depth of measurement effects and the use of lead filters to minimize them, and intensity modulated radiation therapy quality assurance test results compared to that for film. Practical techniques are provided to optimize the accuracy of the system as a dosimetric replacement for film.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...