Related Experiment Videos
A comparison between the old Transcend and the new Transcend series 2000 bracket
1Orthodontic Department, Institute of Dental Surgery, U.M.D.S., London.
Summary
The new Transcend series 2000 bracket shows higher shear bond strength but lower tensile bond strength compared to the original Transcend. Light-cured composite choice significantly impacts bond performance and failure modes.
Area of Science:
- Orthodontic materials science
- Dental adhesive technology
Background:
- The Transcend bracket is a common orthodontic appliance.
- Evaluating new bracket designs and their bonding characteristics is crucial for clinical success.
Purpose of the Study:
- To compare the shear/peel and tensile/peel bond strengths of the original Transcend and the new Transcend series 2000 brackets.
- To assess the influence of two light-cured composites (Prismafil and Heliosit) on bond strength and failure modes.
Main Methods:
- Determining shear/peel and tensile/peel bond strengths of two Transcend bracket types.
- Utilizing two light-cured composites: Prismafil (high filler) and Heliosit (low filler).
- Recording predominant failure sites for each bracket-composite combination and testing mode.
Main Results:
- Transcend series 2000 exhibited higher shear/peel bond strengths than the original Transcend.
- Transcend series 2000 showed significantly weaker tensile/peel bond strengths compared to the original Transcend.
- Heliosit generally yielded higher bond strengths than Prismafil, with fewer enamel fractures observed with Transcend series 2000.
Conclusions:
- The Transcend series 2000 bracket demonstrates superior shear bond strength, necessitating careful debonding in tensile/peel mode.
- Bonding agent selection (Heliosit vs. Prismafil) influences bond strength and potential for enamel damage.
- Clinical recommendations favor tensile/peel debonding for Transcend series 2000 to prevent enamel damage.