Jan Sochman1, Jan H Peregrin, Miloslav Rocek
1Intensive Care Unit, Clinic of Cardiology, Institute for Clinical and Experimental Medicine, Vídenská 1958/9, 140 21 Prague 4, Czech Republic. jan.sochman@medicon.cz
Purpose:
To evaluate the feasibility of one-step implantation of a new type of stent-based mechanical aortic disc valve prosthesis (MADVP) above and across the native aortic valve and its short-term function in swine with both functional and dysfunctional native valves.
Methods:
The MADVP consisted of a folding disc valve made of silicone elastomer attached to either a nitinol Z-stent (Z model) or a nitinol cross-braided stent (SX model). Implantation of 10 MADVPs (6 Z and 4 SX models) was attempted in 10 swine: 4 (2 Z and 2 SX models) with a functional native valve and 6 (4 Z and 2 SX models) with aortic regurgitation induced either by intentional valve injury or by MADVP placement across the native valve. MADVP function was observed for up to 3 hr after implantation.
Results:
MADVP implantation was successful in 9 swine. One animal died of induced massive regurgitation prior to implantation. Four MADVPs implanted above functioning native valves exhibited good function. In 5 swine with regurgitation, MADVP implantation corrected the induced native valve dysfunction and the device's continuous good function was observed in 4 animals. One MADVP (SX model) placed across native valve gradually migrated into the left ventricle.
Conclusion:
The tested MADVP can be implanted above and across the native valve in a one-step procedure and can replace the function of the regurgitating native valve. Further technical development and testing are warranted, preferably with a manufactured MADVP.