Related Experiment Video
Updated: Aug 14, 2026

Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
Published on: January 7, 2016
A nationwide attempt to standardize growth hormone assays
Toshiaki Tanaka1, Katsuhiko Tachibana, Akira Shimatsu
1The GH and Its Related Factors Study Committee, The Foundation for Growth Science, Tokyo, Japan. tanaka-t@ncchd.go.jp
Abstract:
The Growth Hormone (GH) and Its Related Factors Study Committee of the Foundation for Growth Science, Japan, has been conducting a quality control study for 15 years to improve the equality of diagnosis of GH deficiency. It found that the greatest differences in measured GH values were due to the different potencies of the kit standards, which were primarily adjusted to WHO standards for human GH of pituitary origin. With the collaboration of kit makers and the Study Group of Hypothalamo-Pituitary Disorders of the Ministry of Health, Labor and Welfare, all GH kits in Japan have begun using the same recombinant human GH standard since April 2005. As a result the diagnostic cut-off peak GH has changed from 10 to 6 ng/ml.