Magnetic resonance imaging of the body trunk using a single-slab, 3-dimensional, T2-weighted turbo-spin-echo sequence
Matthias Philipp Lichy1, Beate M Wietek, John P Mugler
1Department of Diagnostic Radiology, University of Tübingen, Tübingen, Germany. matthias.lichy@med.uni-tuebingen.de
Purpose:
The authors conducted a clinical evaluation of single-slab, 3-dimensional, T2-weighted turbo-spin-echo (TSE) with high sampling efficiency (SPACE) for high isotropic body imaging with large field-of-view (FoV).
Materials And Methods:
Fifty patients were examined in clinical routine with SPACE (regions of interest: pelvis n=30, lower spine n=12, upper spine n=6, extremities n=4) at 1.5 T. For achieving a high sampling efficiency, parallel imaging, high turbofactor, and magnetization restore pulses were used. In contrast to a conventional TSE imaging technique with constant flip angle refocusing, the refocusing pulse train of the SPACE sequence consists of variable flip angle radiofrequency pulses along the echo train.
Results:
Signal-to-noise ratio and contrast-to-noise ratio of SPACE images were of sufficient diagnostic value. The possibility of image reconstruction in multiple planes was of clinical relevance in all cases and simplified data analysis.
Conclusion:
The achievement of 3-dimensional, T2-weighted TSE magnetic resonance imaging with isotropic and high spatial resolution and interactive 3-dimensional visualization essentially improve the diagnostic potential of magnetic resonance imaging.