Viscoelasticity of pediatric blood and its implications for the testing of a pulsatile pediatric blood pump

Jennifer A Long1, Akif Undar, Keefe B Manning

  • 1Department of Bioengineering, The Pennsylvania State University, University Park, PA 16802, USA.

ASAIO Journal (American Society for Artificial Internal Organs : 1992)
|December 3, 2005
PubMed

Insights

Pediatric blood viscosity and elasticity were measured, revealing similar viscoelastic properties to adult blood. These findings inform the development of pediatric blood pumps using viscoelastic analogs.

Area of Science:

  • Biomedical Engineering
  • Hemodynamics
  • Rheology

Background:

  • Blood rheology, including viscosity and elasticity, is crucial for understanding blood flow, particularly in medical devices like pulsatile blood pumps.
  • Previous research has focused on blood viscosity, but the elasticity of pediatric blood remains largely uncharacterized, hindering device design.
  • The viscoelastic properties of blood significantly impact flow dynamics within small blood pumps, affecting filling characteristics and flow separation.

Purpose of the Study:

  • To measure and characterize the viscosity and elasticity of pediatric blood across a range of hematocrit values.
  • To compare the viscoelastic properties of pediatric blood with adult blood.
  • To develop pediatric blood analog solutions based on measured properties for use in blood pump testing.

Main Methods:

  • Utilized an oscillatory rheometer to measure blood viscosity and elasticity against shear rate.
  • Tested pediatric blood samples with hematocrit levels ranging from 19% to 56%.
  • Statistically analyzed data to compare pediatric and adult blood viscoelasticity at constant hematocrit.

Main Results:

  • Presented comprehensive data on pediatric blood viscosity and elasticity as a function of shear rate and hematocrit.
  • Found minimal influence of patient age on blood viscoelasticity.
  • Demonstrated statistically similar viscoelastic properties between adult and pediatric blood when compared at constant hematocrit.

Conclusions:

  • Pediatric blood exhibits viscoelastic properties comparable to adult blood.
  • Developed validated blood analog solutions for pediatric hematocrit levels.
  • These findings are essential for optimizing the design and performance of pediatric blood pumps by incorporating realistic viscoelastic fluid behavior.