Published on: November 9, 2018
[Influence of photon scattering on the quantification of relative changes in longitudinal brain PET studies with
P Aguiar Fernández1, C Falcón Falcón, C Crespo Vázquez
1Laboratori de Biofísica i Bioenginyeria, Facultat de Medicina, Universitat de Barcelona, Barcelona.
Aim:
To study the effect of photon scattering on the quantification of relative changes of 18F-FDG uptake in longitudinal brain PET studies.
Materials And Methods:
Two studies from a numerical Zubal phantom were simulated. One of these was a basal reference study and the other was an activated study showing an increase or decrease in the uptake in a region of the anterior cingulated cortex. SimSET Monte Carlo code was used to simulate PET sinograms. Primary photons, which did not undergo interactions, and scattered photons, which underwent one or more interactions, were stored in separate files to assess the effect of scattering. Reconstruction was carried out using an iterative algorithm based on ordered subsets of projections (OSEM-2D). The relative changes in uptake were calculated from images reconstructed with all the photons (primary and scattered) and from images reconstructed with only primary photons.
Results:
A linear relationship between the calculated and theoretical values was obtained both for the images reconstructed with all the photons and for those reconstructed with primary photons. Our findings show a relative change recovery of 79% +/- 0.4% for all photons and 91% +/- 0.5% for primary photons only.
Conclusions:
Our results highlight subestimation of relative changes of 12% +/- 0.7% when scattered photons are used. Thus the importance of correcting this degradation in order to improve quantification is shown.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET