A multi-center quality control study of different CA15-3 immunoassays
Walter Hubl1, Thomas Demant, Steffen Albrecht
1Krankenhaus Dresden-Friedrichstadt, Institut für Klinische Chemie und Labormedizin, Dresden, Germany.
Clinical Laboratory
|December 7, 2005
Summary
This study assessed the quality control of CA15-3 tumor marker tests across multiple labs. Results show significant improvements in assay precision over 17 years, highlighting the value of standardized controls for reliable cancer diagnostics.
Area of Science:
- Clinical Chemistry
- Tumor Marker Analysis
- Quality Control in Diagnostics
Background:
- CA15-3 is a crucial tumor marker for monitoring breast cancer.
- Inter-laboratory variability and long-term assay stability are significant challenges in CA15-3 testing.
- Standardized quality control materials are essential for ensuring reliable diagnostic results.
Purpose of the Study:
- To evaluate the long-term performance and precision of CA15-3 assays using a multi-center quality control study.
- To assess the utility of liquid BIOREF CA15-3 control sera for inter-laboratory quality control.
- To investigate potential sources of variability in CA15-3 determinations over time.
Main Methods:
- A multi-center study involving 17 laboratories measuring CA15-3 using seven different assays on automated immunoanalyzers.
- Utilization of BIOREF low, medium, and high level control sera for quantitative analysis.
- Calculation of inter-assay imprecision (coefficient of variation) and assay linearity using concentration ratios.
Main Results:
- Mean CA15-3 values for control sera were determined: 19.3 ± 2.7 kU/l (low), 75.2 ± 11.4 kU/l (medium), and 162.9 ± 37.1 kU/l (high).
- Inter-assay coefficients of variation ranged from 2.9% to 15.5%, indicating varying levels of precision across methods and laboratories.
- Assay linearity was confirmed, and long-term monitoring revealed variability linked to calibration changes, but overall precision has significantly improved compared to data from 17 years prior.
Conclusions:
- Test-independent reference materials like BIOREF sera are effective for CA15-3 quality control.
- These controls enable laboratories to monitor long-term assay performance and detect calibration drift.
- The study demonstrates substantial advancements in CA15-3 assay precision over the past two decades, enhancing diagnostic reliability.