Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts
Published on: April 14, 2023
High-level microwave disinfection of dental gypsum casts
Einar Berg1, Oyunn Nielsen, Nils Skaug
1Faculty of Dentistry, Department of Oral Sciences-Prosthodontics, University of Bergen, Norway. einar.berg@odont.uib.no
Purpose:
To test whether microwave oven irradiation can disinfect gypsum casts in compliance with current disinfection requirements, and to determine whether this procedure would be as effective as a validated method of chemical disinfection of impressions.
Materials And Methods:
In 2 in vitro experiments, samples of 5 irreversible hydrocolloid impressions of a disinfected acrylic resin model were contaminated with suspensions of recommended test organisms Staphylococcus aureus and Pseudomonas aeruginosa, respectively. In 1 in vivo experiment, 10 impressions were made of 10 subjects. All impressions were poured and the resulting casts cut in 2 halves. One half of each cast was exposed to 5 minutes of irradiation in a microwave oven at 2,450 MHz and 900 W. The other halves were left untreated as controls. In a second in vivo experiment, 10 impressions were disinfected by immersion in a 0.07% solution of NaOCl at pH of about 10 for 3 minutes, and then poured. All casts were incubated aerobically in Bacto tryptic soy broth at 37 degrees C for 6 hours and assessed for bacterial growth by counting colony-forming units per milliliter (cfu/mL) of the culture. The results were analyzed using the Wilcoxon signed-rank test.
Results:
Untreated gypsum casts showed cfu/mL counts with a median log value of 6, while microwave-irradiated ones had median cfu/mL counts of 0. Casts poured from chemically disinfected impressions demonstrated cfu/mL counts with a median log value of 4.
Conclusion:
Under the described conditions, microwave-irradiated gypsum casts satisfy current disinfection requirements, but gypsum casts poured from chemically disinfected impressions do not.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Chemical Agents for Microbial Control