Improving hollow fiber dialyzer efficiency with a recirculating dialysate system II: comparison against two-chamber
Manuel Prado1, Laura M Roa, Alfonso Palma
1Biomedical Engineering Group, University of Seville, Seville, Spain. mpradovelasco@ieee.org
Abstract:
The theoretical basis of the nonregenerated recirculating dialysate system (RDS) was derived in Part I of this work [M. Prado, L. M. Roa, A. Palma, and J. A. Milán, Ann. Biomed. Eng. (2005)]. This system pursues the maximization of the clearance of hollow fiber dialyzers whose performance is controlled by diffusion, as occurred in standard hemodialysis. In this second part we perform a comparison by digital simulation of the RDS against three well-known two-chamber dialysis systems. As a major outcome, the efficiency of the RDS increased by a factor of five-eight with respect to the efficiency of a single dialyzer operating with a number of transfer units equal to 0.1, that is when the diffusive mass-transfer of the dialyzer is exhausted. Present low-flux dialyzers do not take advantage of the full potential of this technique, but the functional domain where high-flux and high-area dialyzers operate could be more suitable to exploit this technique. We conclude that RDS can be a competitive efficient technique for optimizing the dialysis efficiency.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis I: Introduction
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy