In vivo band manometry: a new access to band adjustment
Wolfgang Lechner1, Michael Gadenstätter, Ruxandra Ciovica
1Department of Surgery, General Hospital of Krems, Krems, Austria.
Background:
By application of a newly developed device for invasive pressure measurements, we have investigated band adjustments monitored by in vivo intraband pressures. With access to the port of the gastric banding device, pressures can be recorded inside the band system at rest and during bolus application with different adjustments of the band.
Methods:
25 patients (mean age 38.7, mean BMI 45.1, 80% women) had intraband pressure measurements at the first band adjustment 8.2 weeks (range 6 to 17) postoperatively. For this purpose, we adapted a pressure monitoring system with the TruWave disposable pressure transducer of Edwards. All patients underwent gastric banding using the Swedish adjustable gastric band (SAGB) by the pars flaccida technique.
Results:
In vivo intraband pressures differ from ex vivo intraband pressures. With increasing fill volume in vivo measurements show increasingly higher pressures than ex vivo measurements. This difference can mainly be attributed to the influence of the enclosed tissue. The in vivo intraband pressures correlate with the amount of outflow obstruction.
Conclusion:
Intraband pressure measurement is an encouraging new access to gastric banding. It appears to be a feasible method to control band adjustment without need for x-ray studies in low pressure bands. We expect physiologically exact adjustments to achieve good weight loss and to prevent esophageal problems in the long term.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)