A Novel Surgical Technique in a Sheep Model for Suburethral Graft Implantation
Published on: June 20, 2025
Novel everting urologic access sheath: decreased axial forces during insertion
Jonathan N Rubenstein1, Maurice Garcia, Affonso H L A Camargo
1Department of Urology, University of California, San Francisco, San Francisco, California 94143-1611, USA.
Journal of Endourology
|December 20, 2005
Summary
A novel everting urethral sheath significantly reduced forces during urologic instrument insertion. This technology may decrease patient discomfort and trauma during lower urinary tract endoscopy.
Area of Science:
- Urology
- Biomedical Engineering
- Medical Device Innovation
Background:
- Urologic instrument advancement generates significant axial forces, causing patient discomfort and potential trauma.
- These forces can lead to mucosal damage, dysuria, hematuria, infection, and strictures.
- An everting urethral sheath may mitigate these adverse effects.
Purpose of the Study:
- To evaluate the efficacy of a novel everting urethral sheath in reducing axial forces during urologic procedures.
- To compare the force reduction capabilities of the everting sheath against traditional methods.
Main Methods:
- Two urethral luminal models (7-cm long, 5-mm diameter) were created, one with an artificial stricture.
- Axial forces were measured during the advancement of a novel everting access sheath (Cystoglide) and compared to a cystoscope and a traditional access sheath.
- Tests were conducted with and without lubrication and with and without an artificial stricture.
Main Results:
- The everting sheath generated significantly lower mean advancement forces compared to both the cystoscope (P<0.01) and the traditional access sheath (P<0.01).
- This force reduction was consistent in models with and without artificial strictures (P<0.01).
- Lubrication also did not significantly alter the force reduction benefit of the everting sheath (P<0.01).
Conclusions:
- The novel everting introduction sheath substantially decreased axial forces on the urethral luminal wall.
- Clinical application of this technology holds potential for reducing discomfort and complications in lower urinary tract endoscopy.