Related Experiment Video
Updated: Aug 14, 2026

Measurement of Tumor T2* Relaxation Times after Iron Oxide Nanoparticle Administration
Published on: May 19, 2023
R2 relaxometry with MRI for the quantification of tissue iron overload in beta-thalassemic patients
Efthymia Alexopoulou1, Fotini Stripeli, Panagiotis Baras
1Second Department of Radiology, National and Kapodistrian University of Athens, Athens, Greece.
Purpose:
To evaluate the usefulness of a time-efficient MRI method for the quantitative determination of tissue iron in the liver and heart of beta-thalassemic patients using spin-spin relaxation rate, R2, measurements.
Materials And Methods:
Images were obtained at 1.5 T from aqueous Gd-DTPA solutions (0.106-8 mM) and from the liver and heart of 46 beta-thalassemic patients and 10 controls. The imaging sequence used was a respiratory-triggered 16-echo Carr-Purcell-Meiboom-Gill (CPMG) spin-echo (SE) pulse sequence (TR = 2000 msec, TE(min) = 5 msec, echo spacing (ES) = 5 msec, matrix = 192 x 256, slice thickness = 10 mm). Liver iron concentration (LIC) measurements were obtained for 22 patients through biopsy specimens excised from the relevant liver segment. Biopsy specimens were also evaluated regarding iron grade and fibrosis. Serum ferritin (SF) measurements were obtained in all patients.
Results:
A statistically significant difference was found between patients and healthy controls in mean liver (P < 0.004) and myocardium (P < 0.004) R2 values. The R2 values correlated well with Gd DTPA concentration (r = 0.996, P < 0.0001) and LIC (r = 0.874, P < 0.0001). A less significant relationship (r = 0.791, P < 0.0001) was found between LIC measurements and SF levels. R2 measurements appear to be significantly affected (P = 0.04) by different degrees of hepatic fibrosis. The patients' liver R2 values did not correlate with myocardial R2 values (r = 0.038, P < 0.21).
Conclusion:
Tissue iron deposition in beta-thalassemic patients may be adequately quantified using R2 measurements obtained with a 16-echo MRI sequence with short ES (5 msec), even in patients with a relatively increased iron burden.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI