Background:
We developed a non-invasive, continuous, high-resolution method of measuring carboxyhaemoglobin fraction (COHb%) using expiratory gas analysis (EGA). We assessed whether application of EGA to carboxyhaemoglobin dilution provides red cell volume (RCV) measurement with accuracy equivalent to that of CO-haemoximetry, with a smaller infusion volume of carbon-monoxide-saturated autologous blood (COB). Method. We assessed the agreement between repeated COHb% measurements by EGA and simultaneous measurement by CO-haemoximetry, using Bland and Altman plot, in healthy subjects and patients with artificially controlled ventilation and no radiological evidence of pulmonary oedema or atelectasis. We assessed the agreement between RCV measurements by EGA with infusion of 20 ml of COB (RCVEGA) and RCV measurements by CO-haemoximetry with infusion of 100 ml of COB (RCVHEM), in healthy subjects.
Results:
The 'limits of agreement' between COHb% measurement by EGA (1 min average) and CO-haemoximetry were -0.09 and 0.08% in healthy subjects, and -0.11 and 0.09% in patients. Given the resolution of CO-haemoximetry (0.1%), the accuracy of EGA was equivalent to or greater than that of CO-haemoximetry. The 'limits of agreement' between RCVEGA and RCVHEM were -0.14 and 0.15 litre. Given the average resolution of RCVHEM (0.14 litre), the accuracy of RCVEGA was equivalent to that of RCVHEM.
Conclusion:
EGA provided non-invasive, accurate, continuous, high-resolution COHb% measurements. Applying EGA to carboxyhaemoglobin dilution, we achieved RCV measurements with accuracy equivalent to that of CO-haemoximetry, with one-fifth of the COB infusion volume. However, clinical application of the method is limited to patients with no radiological evidence of pulmonary oedema or atelectasis.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...