The performance of commercial photodiodes for dosimetry in mammography
E Batista1, H J Khoury, F A Melo
1Departamento de Energia Nuclear, Universidade Federal de Pernambuco, Cidade Universitária, 50740-540 Recife, PE, Brazil.
Abstract:
This paper reports the results of an investigation carried out to determine the suitability of using a commercially available photodiode for dosimetry in mammography thereby providing a low cost dosemeter with a direct reading of the radiation dose. A mammographic X-ray generator (Siemens Mammomat 1000) with tube potential in the 23-30 kV range and a constant potential X-ray generator (Pantak) with a W/Mo anode/filter combination were used in this study. The results showed that the photodiode response is highly linear within mammographic dose ranges and that the energy dependence was <3% at tube potentials of 25-30 kV. Good agreement was observed between the incident air kerma measured with both the photodiode and the ionization chamber. These results show the viability of using the photodiode as a dosemeter system in mammography.