Optimal scanning protocol of multislice CT virtual intravascular endoscopy in pre-aortic stent grafting: in vitro
Zhonghua Sun1, Charlene Ferris
1Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University of Technology, GPO Box U 1987, Perth, WA 6845, Australia. z.sun@curtin.edu.au
Objective:
To investigate the optimal scanning protocol for multislice computed tomography angiography (MSCT) in pre-aortic stent grafting observed with virtual intravascular endoscopy (VIE).
Materials And Methods:
The study was performed on a human abdominal aortic phantom which was housed in a perspex container, filled with contrast medium having CT attenuation similar to that used in the patient's abdominal CT scan. A series of scans were performed on a four-slice multislice CT scanner with the scanning protocols as follows: section thickness of 1.3, 3.2 and 6.5 mm, pitch value of 0.875, 1.25 and 1.75 with reconstruction intervals of 50% overlap. The degree of stair-step artifacts was measured at three different locations, superior mesenteric artery (SMA), renal ostium and the normal abdominal aorta. Standard deviation (S.D.) of the signal intensity measured on surface shaded images was used to determine the image quality. Radiation dose was also recorded in each scanning protocol.
Results:
The VIE images showed that image quality was not dependent on pitch and section thickness in the visualization of renal ostium and SMA, whereas it was dependent on these two factors at the level of the normal aorta (p<0.05). It was noticed that when section thickness reached 6.5 mm the SMA and renal ostia became distorted. Radiation dose measured in 1.3 mm protocols was significantly higher that those measured in other section thicknesses (p<0.05).
Conclusion:
The scanning protocol of section thickness 3.2 mm, pitch 1.25 with a reconstruction interval of 1.6 mm was recommended as it allows optimal visualization of VIE images of aortic ostia, generation of fewer artifacts and less radiation dose.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...