Related Experiment Video
Updated: Aug 13, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Technical note: gold marker implants and high-frequency jet ventilation for stereotactic, single-dose irradiation of
1St. Marien-Krankenhaus, Klinik für Radio-Onkologie, Kampenstrasse 51, D-57072 Siegen, Germany. P.H.Fritz@t-online.de
Abstract:
With reference to radiosurgery of the liver, we describe techniques designed to solve the methodological problem of striking targets subject to respiratory motion with the necessary precision. Implanting a gold marker in the vicinity of the liver tumor was the first step in ensuring the reproducibility of the isocenter's position. An 18-karat gold rod measuring 1.9 x 3 mm was implanted approximately 2 cm from the edge of the tumor as this was displayed in the spiral, thin-slice CT with contrast media. Both the implantation of the marker and the required, CT-controlled biopsy of the liver tumor can be achieved simultaneously with the same puncture needle. The efficiency of high-frequency jet ventilation (HFJV) in neutralizing the targeted organ's respiratory motion during stereotactic single-dose irradiation was evaluated. The procedure was carried out on ten patients without any complications. In the time between treatment planning and irradiation (3 days), no significant marker migration was observable. In all cases, the gold marker (volume: 7.5 mm(3)) was readily observable in the treatment beam using portal imaging. HFJV provided reliable immobilization. The liver motion in each anesthetized patient was limited to under 3.0 mm in all directions. Thus, the correct field settings and target reproducibility were able to be analyzed and documented during the irradiation. The combination of marker and HFJV enables the determination of stereotactic coordinates directly related to the liver itself and, in this way, stereotactic radiation treatment of liver tumors is freed from the uncertainties involved in orientation to bony landmarks, in respiratory motion, and in changes of position in the stereotactic body frame. The method is feasible and can improve the accuracy of stereotactic body radiation therapy.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
10:25A Dorsal Skinfold Window Chamber Tumor Mouse Model for Combined Intravital Microscopy and Magnetic Resonance Imaging in Translational Cancer Research
Published on: April 12, 2024