Detectability decreases with off-normal viewing in medical liquid crystal displays
Aldo Badano1, Brandon D Gallas
1NIBIB/CDRH Laboratory for the Assessment of Medical Imaging, Center for Devices and Radiological Health, Food and Drug Administration, Rockville, MD 20857, USA. aldo.badano@fda.hhs.gov
Rationale And Objectives:
To quantify the reduction in detection performance of subtle signals at off-normal viewing directions in medical active-matrix liquid crystal displays (AMLCDs).
Materials And Methods:
Fifty synthetic image pairs per viewing condition (a total of 350) were used in a two-alternative forced-choice experiment in which 11 trained observers viewed images at 0, 30, and 45 degrees from the display normal, along the diagonal axis of a 5 million pixel in-plane switching monochrome AMLCD. The images were generated using white-noise backgrounds. A Gaussian signal was added to the signal-present set with three different signal amplitudes (4, 8, and 12 gray levels in a 10-bit scale).
Results:
The average percent correct achieved for a signal of 4 gray levels was 79.6 (95% confidence intervals based on reader and case variability: 71.6-86.9), 63.4 (CI 56.0-71.3), and 55.3 (CI 48.4-62.0), for 0, 30 and 45 degrees from the display normal, respectively. When the signal amplitude was increased by a factor of two, the performance was 76.9 and 57.0 for 30 and 45 degrees, respectively, and 95.3 and 85.3 when the amplitude was increased by a factor of three. The observers took on average about twice as long and as much as seven times as long to reach decisions in off-normal viewing.
Conclusions:
Off-normal viewing of diagnostic images in AMLCDs significantly reduces the detection of low-contrast abnormalities. Increased off-normal signal amplitudes were required to regain the detection performance measured for normal viewing. We observed this decrease in detection performance for off-normal viewing even when measured decision times were about twice as long as for normal viewing.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...