Modified nasal cannula for simultaneous oxygen delivery and end-tidal CO2 monitoring during spontaneous breathing
1Department of Anesthesiology and Pain Medicine, Gifu University Graduate School of Medicine, Gifu, Japan.
Background And Objectives:
Supplemental oxygen is commonly given via nasal cannulae in spontaneously breathing patients. Our modified nasal cannula with a clamp between the nasal prongs can provide O2 via one nostril and CO2 can be sampled through the other one. We have studied whether this cannula can provide oxygenation similar to a standard cannula without affecting end-tidal CO2 monitoring.
Methods:
Eighty-six patients were studied during spinal anaesthesia and sedation. In 15 patients, arterial blood was sampled while O2 was delivered at flow rates of 0, 2 and 4 L min(-1), with or without clamping between the prongs of our modified nasal cannula. In the remaining 71 patients, arterial O2 was measured while using our modified nasal cannula with the clamp applied. End-tidal CO2 was recorded on a capnograph and the correlation between end-tidal and arterial values with our modified nasal cannula was investigated.
Results:
No end-tidal CO2 waveforms were found with oxygen flow greater than 2L min(-1) without clamping between the prongs. With clamping there was a significant correlation (r = 0.83) between arterial and end-tidal CO2. A Bland-Altman analysis revealed a bias of 0.49 kPa with precision of +/-0.76 kPa. Arterial oxygenation was not affected by our modified nasal prongs with clamp as compared to the standard cannula.
Conclusion:
Our modified nasal cannula can provide continuous monitoring of end-tidal CO2 without affecting oxygen delivery in sedated, spontaneously breathing patients.
More Related Videos
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy