Related Experiment Video
Updated: Jul 12, 2026

Endoscopic Endonasal Trans-sphenoidal Approach: Minimally Invasive Surgery for Pituitary Adenomas
Published on: January 17, 2018
Irrigation-suction straw sheath system for a rigid endoscope during endonasal endoscopic pituitary surgery
S Kubo1, T Kikawada, H Hasegawa
1Department of Neurosurgery, Tominaga Hospital, 1-4-48 Minato-machi, Naniwa-ku, Osaka 556-0017, Japan. sig-kubo@momo.so-net.ne.jp
Objective:
One of the most problematic conditions during endonasal endoscopic pituitary surgery is blurred endoscopic vision caused by blood or bone dust. To address this problem we used a new device, the irrigation-suction straw sheath system originally developed for endoscopic sinus surgery in rhinology, for the endoscopic pituitary surgery.
Methods:
The irrigation-suction straw system (K-endosheath; Koken Co., Tokyo, Japan) consists of a flattened disposable straw catheter which can be attached on a rigid endoscope. When the endoscope is inserted into the oblong catheter, two gaps remain on either side of the endoscope. Through each of these gaps irrigation and continuous suction are performed respectively.
Results:
Clear vision could be immediately obtained by pressing the button to release the irrigation water onto the endoscope's front lens. Even targeted irrigation during drilling was possible. Above all, continuous suction was useful to clear the operative field of blood, smoke, or bone dust, ensuring the safety and uninterrupted progress of the procedure.
Conclusion:
This irrigation-suction system has proved to be very useful and easy to use for endonasal endoscopic pituitary surgery. Further development of such new devices and instruments specialized for endonasal endoscopic pituitary surgery will promote the increased use of this minimally invasive technique in neurosurgery.
Related Concept Videos
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Suctioning the Nasopharyngeal Airway
Equipment Required
Tracheostomy Suctioning II: Procedure
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...