Thermal threshold testing for the assessment of small fibre dysfunction: normal values and reproducibility
B Bravenboer1, P S van Dam, J Hop
1University Hospital Utrecht, The Netherlands.
Summary
This study establishes normal thermal perception thresholds for warmth and cold in healthy adults using the Thermal Threshold Tester. Age significantly impacts cold perception in the feet and warm perception in the hands.
Area of Science:
- Neurology
- Diabetology
- Sensory Physiology
Background:
- Diabetic mellitus frequently leads to small and large fiber neuropathy.
- Accurate assessment of thermal perception thresholds is crucial for diagnosing neuropathy.
- Establishing normative data is essential for interpreting patient results.
Purpose of the Study:
- To determine normal values for warm and cold thermal perception thresholds.
- To assess the influence of age on thermal perception in healthy individuals.
- To evaluate the repeatability of thermal threshold measurements.
Main Methods:
- Utilized the Thermal Threshold Tester device to generate controlled thermal stimuli.
- Measured warm and cold thresholds using a forced-choice method with an up-and-down transform rule.
- Assessed 80 healthy subjects across four age groups (25-65 years) at the wrist and foot.
- Evaluated repeatability in 39 diabetic patients without neuropathy.
Main Results:
- A significant age-related increase in warm thermal threshold was observed in the hand.
- A significant age-related increase in cold thermal threshold was observed in the foot.
- Thermal thresholds in the hand (cold) and foot (warm) did not show significant age-related differences.
- Good repeatability was found for hand measurements, but poor repeatability for foot measurements.
Conclusions:
- Normal thermal perception thresholds were established for healthy adults.
- Age is a significant factor influencing thermal perception, particularly cold sensation in the feet and warm sensation in the hands.
- The Thermal Threshold Tester demonstrates good repeatability for hand measurements but requires further validation for foot measurements.