Development of a removable multi-column-switching system for gas chromatography
1Schering AG, Global Chemical Development, In-Process Control Department, 13342 Berlin, Germany. dirk.hinz@schering.de
A new removable column-switching device for gas chromatography (GC) was developed. This system efficiently screens multiple GC columns, avoiding adsorption and degradation for reliable analysis.
Area of Science:
- Analytical Chemistry
- Chromatography
Background:
- Gas chromatography (GC) method development often requires screening multiple stationary phases.
- Existing methods for column screening can be time-consuming and labor-intensive.
Purpose of the Study:
- To develop a removable column-switching device for commercial gas chromatographs.
- To address challenges including high temperature stability, gas flow through standby columns, catalytically active surfaces, valve cooling trap effects, and limited oven space.
Main Methods:
- Development of a novel column-switching system integrated into a commercial GC.
- Investigation of the system's performance using dead volume (tailing factor), adsorption (Grob test), and degradation (Endrin test) analyses.
Main Results:
- The developed GC column-switching system demonstrated comparable results to analyses without column switching.
- The system effectively avoided issues such as adsorption and decomposition.
- The device successfully tested the suitability of up to eight separating columns for unknown samples.
Conclusions:
- The column-switching system enhances efficiency in selecting or screening stationary phases for GC method development.
- This innovation streamlines the process of optimizing GC methods for diverse analytical challenges.
Related Concept Videos
Principles Of Column Chromatography
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Instrumentation