Related Experiment Video
Updated: Aug 13, 2026

Amplification of Escherichia coli in a Continuous-Flow-PCR Microfluidic Chip and Its Detection with a Capillary Electrophoresis System
Published on: November 21, 2023
Toward million-fold sensitivity enhancement by sweeping in capillary electrophoresis combined with thermal lens
Fumihiko Kitagawa1, Takashi Tsuneka, Yoshihiro Akimoto
1Department of Material Chemistry, Graduate School of Engineering, Kyoto University, Katsura, Kyoto 615-8510, Japan. fkitagawa@mbox.kudpc.kyoto-u.ac.jp
Abstract:
This paper reports a thermal lens microscope (TLM) detection coupled with capillary electrophoresis (CE) by using an interface chip (IFChip) to achieve highly sensitive detection with high reproducibility. Fused silica capillaries with an inner diameter of 50 microm were directly connected to a microchannel on the IFChip. In comparison with an on-capillary detection method in CE-TLM, ca. 10-fold improvements in the reproducibility for peak height were obtained by using IFChips. The detection limit of an azo dye was estimated to be 3.6 x 10(-7)M (100 ppb), which was above 100-times lower than that of conventional absorbance detection. Toward further improvement of the detectability for nonfluorescent compounds, on-line sample preconcentration by sweeping was applied to the CE-TLM using the IFChip. Due to the sweeping effect, 3900000-fold increase in the sensitivity was successfully achieved.
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...