Related Experiment Video
Updated: Aug 13, 2026

A Novel Surgical Technique in a Sheep Model for Suburethral Graft Implantation
Published on: June 20, 2025
Five cases of tape erosion after transobturator surgery for urinary incontinence
Magali Robert1, Magnus Murphy, Colin Birch
1Division of Pelvic Floor Disorders and Reconstructive Surgery, Department of Obstetrics and Gynaecology, University of Calgary, Calgary, Alberta, Canada.
Background:
Before introducing the transobturator tape into our practice we undertook 52 transobturator tape procedures. The transobturator tape procedures were undertaken with an "outside-in" approach, using nonwoven polypropylene mesh with average pore size of 50 mum.
Cases:
Five cases of vaginal erosions have been identified, 1 complicated by a groin abscess. All cases required further procedures to trim (n = 3), resect (n = 1) or remove (n = 1) the tape. One woman had a tension-free vaginal tape procedure. To date, 3 women remain incontinent.
Conclusion:
Possible reasons for the complications include 1) surgical inexperience (unlikely, given that we have undertaken more than 2000 tension-free vaginal tape procedures without similar complication rates); 2) inherent susceptibility of the "hammock" position of the transobturator tape; or 3) the nonwoven polypropylene tape with mesh size of 50 microm itself may predispose to erosion or abscess.