Related Experiment Video
Updated: Aug 11, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron
Irmgard Griessbach1, Markus Lapp, Jörg Bohsung
1Clinic for Radiotherapy, University of Magdeburg, Leipziger Str. 44, 39120 Magdeburg, Germany. irmgard.griessbach@medizin.uni-magdeburg.de
Abstract:
Shielded p-silicon diodes, frequently applied in general photon-beam dosimetry, show certain imperfections when applied in the small photon fields occurring in stereotactic or intensity modulated radiotherapy (IMRT), in electron beams and in the buildup region of photon beam dose distributions. Using as a study object the shielded p-silicon diode PTW 60008, well known for its reliable performance in general photon dosimetry, we have identified these imperfections as effects of electron scattering at the metallic parts of the shielding. In order to overcome these difficulties a new, unshielded diode PTW 60012 has been designed and manufactured by PTW Freiburg. By comparison with reference detectors, such as thimble and plane-parallel ionization chambers and a diamond detector, we could show the absence of these imperfections. An excellent performance of the new unshielded diode for the special dosimetric tasks in small photon fields, electron beams and build-up regions of photon beams has been observed. The new diode also has an improved angular response. However, due to its over-response to low-energy scattered photons, its recommended range of use does not include output factor measurements in large photon fields, although this effect can be compensated by a thin auxiliary lead shield.
Related Concept Videos
Schottky Barrier Diode
Diode: Forward bias
The behavior of a diode in forward bias...
Biological Effects of Radiation
Diode: Reverse bias