Rationale And Objectives:
To evaluate the correlation between image noise and body weight (BW) or body mass index (BMI) in coronary computed tomography angiography (CTA) as a potential parameter for reducing radiation dose in coronary CTA.
Materials And Methods:
Thirty-six patients who underwent electrocardiogram-gated cardiac CT were analyzed in this study. The patients included 26 men and 10 women with a mean age of 60 years (range 43-79 years). All patients were imaged on a 16-row multidetector CT scanner. Mean value of BW and BMI was 83.5 kg and 28.1, respectively. Image noise was defined as standard deviation (SD) of the attenuation values measured by using 1 cm2 circular region of interest in the ascending aorta at the level of the right main pulmonary artery. The SD values were plotted against BW and BMI. The correlations were examined using a linear regression method. A P value of less than .05 was considered significant.
Results:
The r value of linear regression between noise and BW was 0.90 (P < .001). The r value of linear regression between noise and BMI was 0.74 (P = .015).
Conclusions:
Excellent correlation was observed between noise and BW in coronary CTA. These data may be used as potential parameters for customized radiation dose modification to reduce radiation dose in coronary CT examinations.
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI