Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Analysis of the optical concentration curve to detect access recirculation
L Mercadal1, B Coevoet, M Albadawy
1Nephrology Department, Pitié-Sapétrière Hospital, Paris, France. lucile.mercadal@psl.ap-hop-paris.fr
Kidney International
|March 7, 2006
Summary
Optical blood volume curve spikes during ionic dialysance (ID) measurement indicate access recirculation in dialysis patients. This finding offers a reliable method for detecting recirculation, especially when exceeding 20%.
Area of Science:
- Nephrology
- Medical Devices
- Diagnostic Techniques
Background:
- The optical blood volume curve can exhibit rapid variations (spikes) during ionic dialysance (ID) measurements.
- The cause and significance of these spikes, particularly in relation to access recirculation, require investigation.
Purpose of the Study:
- To determine if spikes in the optical blood volume curve are associated with access recirculation during hemodialysis.
- To assess the diagnostic value of these spikes for detecting and quantifying access recirculation.
Main Methods:
- Investigated the effect of reversed blood line positioning on the blood volume curve to simulate recirculation.
- Systematically monitored for spikes during ID measurements in patients undergoing dialysis.
- Quantified access recirculation using an ultrasound dilution technique in parallel with spike observation.
Main Results:
- Reversing blood lines consistently reproduced spikes in the blood volume curve when recirculation exceeded 20%.
- Patients with access recirculation greater than 20% exhibited spikes during normal ID measurements.
- Spikes were inconsistent for recirculation between 10-20% and absent below 10% or undetectable recirculation.
Conclusions:
- The presence of spikes in the optical blood volume curve during ID measurement is a clear indicator of access recirculation.
- This phenomenon demonstrates 100% specificity when spikes are present throughout the dialysis session for all ID measurements.
- A sensitivity of 100% is achieved for detecting access recirculation greater than 20%.