Shortcomings of protein removal prior to high performance liquid chromatographic analysis-A case study using method
A K Hewavitharana1, C Hyde, R Thomas
1School of Pharmacy, University of Queensland, Brisbane, Qld 4072, Australia. a.hewavitharana@pharmacy.uq.edu.au
Abstract:
During the analytical method development for BAY 11-7082 ((E)-3-[4-methylphenylsulfonyl]-2-propenenitrile), using HPLC-MS-MS and HPLC-UV, we observed that the protein removal process (both ultrafiltration and precipitation method using organic solvents) prior to HPLC brought about a significant reduction in the concentration of this compound. The use of a structurally similar internal standard, BAY 11-7085 ((E)-3-[4-t-butylphenylsulfonyl]-2-propenenitrile), was not effective in compensating for the loss of analyte as the extent of reduction was different to that of the analyte. We present here a systematic investigation of this problem and a new validated method for the determination of BAY 11-7082.