Is the "instant flow" option on the NIOX analyser needed?
1Department of Child Health, Royal Aberdeen Children's Hospital, University of Aberdeen, Foresterhill, Aberdeen AB25 2ZG, United Kingdom.
Medical Engineering & Physics
|March 15, 2006
Summary
The restriction of instant flow (IF) on the NIOX analyser does not impact fraction of nitric oxide in exhaled breath (FE(NO)) measurements in children. However, it significantly reduces the success rate of obtaining these crucial asthma diagnostic readings.
Area of Science:
- Respiratory Medicine
- Pediatric Pulmonology
- Biomedical Engineering
Background:
- Elevated levels of gaseous nitric oxide in exhaled breath are characteristic of asthma.
- Fraction of nitric oxide in exhaled breath (FE(NO)) measurements are vital for asthma diagnosis and management.
- The NIOX chemiluminescence analyser is a common tool for FE(NO) assessment.
Purpose of the Study:
- To investigate the effect of the Instant Flow (IF) restriction feature on the NIOX analyser.
- To determine if IF restriction influences the accuracy of FE(NO) measurements.
- To assess the impact of IF restriction on the success rate of obtaining FE(NO) measurements in children.
Main Methods:
- A study was conducted on 38 children with a median age of 10 years.
- FE(NO) measurements were taken from the same individuals with and without the IF restriction enabled on the NIOX analyser.
- Statistical analysis was performed to compare FE(NO) values and success rates between the two conditions (p=0.002 for success rate).
Main Results:
- FE(NO) values obtained with and without IF restriction were not significantly different within individuals.
- The success rate for obtaining FE(NO) measurements was higher when IF was not restricted (57%).
- Conversely, the success rate was lower when IF was restricted (41%), indicating a significant negative impact.
Conclusions:
- The Instant Flow (IF) restriction feature on the NIOX analyser does not alter the measured FE(NO) values.
- The IF restriction significantly decreases the likelihood of successfully obtaining FE(NO) measurements in pediatric subjects.
- Current findings do not support the use of IF restriction on the NIOX analyser for FE(NO) measurements in children.