Performance evaluation of a computed radiography imaging device using a typical "front side" and novel "dual side"
Kenneth A Fetterly1, Beth A Schueler
1Diagnostic Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA. fetterly.kenneth@mayo.edu
Abstract:
The Fourier-space modulation transfer function (MTF), normalized noise power spectrum (nNPS), and detective quantum efficiency (DQE) of a computed radiography (CR) x-ray imaging device were measured. Two different storage phosphor screens were used in conjunction with a single CR reader (Fuji, Clearview, CS). One of the storage phosphor plates (ST-BD) had a clear backing material which allowed "dual side read" of the latent image from both the "front" and "back" sides of the phosphor. The other phosphor plate had a light occluding backing material, limiting the readout to front side only (ST-VI). The standard RQA-5 beam quality was used. The MTF was measured using a 1 mm thick tungsten edge device. Compared to the ST-55BD phosphor, the ST-VI phosphor was found to have modestly higher MTF at all spatial frequencies. The nNPS(f) and DQE(f) were measured for nominal incident exposure levels ranging from 0.1 to 10 mR. The dual side read phosphor demonstrated superior DQE, especially at low spatial frequencies. At the frequency 0.5 cycles/mm, the DQE values for the 1 mR exposure were 0.36 and 0.21 for the ST-55BD and ST-VI phosphor plates, respectively. The differences between the spatial-frequency dependent DQE of the two plates can be attributed to the increased signal collection efficiency of the dual side read plates and differences in storage phosphor structure noise.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
Determination of Crystal Structures
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET