Validation of TM-2655 oscillometric device for blood pressure measurement
Zhanna D Kobalava1, Yulia V Kotovskaya, Lala A Babaeva
1Department of Cardiology and Clinical Pharmacology, Russian Peoples Friendship University, Moscow, Russia.
Objective:
To perform validation for an arm-type oscillometric TM-2655 device (A&D Company Ltd, Tokyo, Japan) for blood pressure measurement according to the British Hypertension Society protocol.
Methods:
Eighty-five study participants (33 men and 52 women) were included in the study. Mean age was 52.9+/-15.0 years, systolic blood pressure range was 84-208 mmHg and diastolic blood pressure range was 48-120 mmHg. For each participant, three readings of TM-2655 were compared with sequential auscultatory measurements by two trained independent observers. The observers used a calibrated mercury sphygmomanometer and dual stethoscope. The results were graded according to the British Hypertension Society protocol 1993.
Results:
The average difference between mercury sphygmomanometer and TM-2655 readings for systolic blood pressure was -1.0+5.2 mmHg (mean+/-SD) and for diastolic blood pressure -0.9+/-4.7 mmHg. The proportions of values agreeing to within 5, 10 and 15 mmHg were 72.5, 93.7 and 99.6% for systolic blood pressure and 78.8, 96.9 and 100% for diastolic blood pressure between the observers and the device (A/A British Hypertension Society grade).
Conclusions:
The TM-2655 device achieved British Hypertension Society grade A/A and therefore can be recommended for blood pressure measurement in an adult population.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement