Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
Subsecond fluorine-19 MRI of the lung
Ursula Wolf1, Alexander Scholz, Claus Peter Heussel
1Department of Radiology, Johannes Gutenberg University, Mainz, Germany.
Abstract:
Minimal scan times in rapid fluorine-19 MRI using sulfur hexafluoride (SF6) have been on the order of 10 s. Because of the very short T1 relaxation time of SF6 (T1 = 1.65 ms), high receiver bandwidths are necessary to allow for a high number of excitations. Since high bandwidths cause high levels of electronic noise, SNR per acquisition has been too low to further reduce scan time. The purpose of this study was to investigate whether scan times could be reduced using hexafluoroethane (C2F6), a gas with a longer T1 (T1 = 7.9 ms) at a relatively low bandwidth of 488 Hz/pixel. Gradient-echo images were acquired during and after completion of the wash-in of a 70% C2F6- 30% O2 mixture. Peak SNR values of 16 and 7.9 were observed for coronal projection images acquired within 2 s and 260 ms, respectively. These results demonstrate that subsecond imaging is feasible using C2F6.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Magnetic Resonance Imaging