Abstract:
The Association of Official Analytical Chemists (AOAC) test for sporicidal activity of disinfectants (966.04) is used in the United States as the legal criteria for classifying a liquid as a chemical sterilant and also as the indicator of the highest level of disinfectant. This qualitative test contains procedures that may cause inaccurate results. A modified, quantitative version of the AOAC sporicidal test has been developed that uses a specified minimum number of spores (2.0 x 10(sup5)) of Bacillus subtilis or Clostridium sporogenes dried onto porcelain penicylinders. This modified test was validated with three commercial sterilant chemicals tested on three separate groups of spore-labeled cylinders.
Related Concept Videos
Chemical Agents for Microbial Control
Antimicrobial Effectiveness
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...