Intertester reliability of assessing postural sway using the chattecx balance system
Journal of Athletic Training
|September 1, 1995
Summary
This study assessed the reliability of balance testing using the Chattecx Dynamic Balance System. Results showed variable reliability, indicating a need for further research on balance assessment protocols.
Area of Science:
- Biomechanics
- Kinesiology
- Physical Therapy
Background:
- Balance assessment is crucial in physical therapy and sports science.
- The Chattecx Dynamic Balance System offers a method for evaluating both static and dynamic balance.
- Intertester reliability is essential for consistent and dependable clinical assessments.
Purpose of the Study:
- To evaluate the intertester reliability of the Chattecx Dynamic Balance System for static and dynamic balance assessments.
- To determine the consistency of measurements across different testers.
- To identify potential variability in balance testing protocols.
Main Methods:
- Twelve subjects (10 female, 2 male) were tested by two independent testers.
- Ten preprogrammed balance conditions were administered, including double-leg and single-leg stances, with eyes open and closed.
- Balance was quantified by measuring postural sway in centimeters using the Chattecx system.
Main Results:
- Intraclass correlation coefficients (ICCs) ranged from poor to excellent (0.06 to 0.90).
- Standard error of measurement (SEM) values varied from 0.06 to 0.34 across trials.
- Mean postural sway values ranged from 0.28 cm to 1.72 cm for double-leg and 0.65 cm to 1.70 cm for single-leg stances.
Conclusions:
- Significant variability in intertester reliability was observed, suggesting potential inconsistencies in balance assessments.
- Further research is recommended to establish robust intratester and intertester reliability using diverse assessment protocols.
- Standardized protocols are needed to enhance the reliability of dynamic and static balance evaluations.