Related Experiment Video
Updated: Aug 9, 2026

Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
Asymmetric vault of an accommodating intraocular lens
Daniela Jardim1, Barrie Soloway, Christopher Starr
1New York Eye and Ear Infirmary, New York, New York 10022, USA.
Abstract:
We report a case of a new syndrome, late asymmetric vault of the AT45 Crystalens intraocular lens (IOL) (Eyeonics Inc.). The late vault, which occurred in 6 month after implantation of the IOL, caused a decrease in uncorrected distance and near acuities and minus-cylinder axis perpendicular to the haptics. Treatment with neodymium:YAG laser capsulotomy and fibrotic band release was successful.