LET and dose dependence of TLD-100 glow curve after exposure to intermediate-energy ions
G Massillon-Jl1, I Gamboa-deBuen, M E Brandan
1Instituto de Física UNAM, AP 20-364, 01000 DF, Mexico. massillon@fisica.unam.mx
Abstract:
We present results from measurements performed with low fluences (10(5)-10(6) cm(-2)) of 15, 25 and 40 MeV u(-1) carbon, 25 MeV u(-1) oxygen and 40 MeV u(-1) neon ions incident on TLD-100 chips. Dosemeters were arranged individually or in stacks in front of the beam, allowing the study of various linear energy transfer (LET) values simultaneously. The thermoluminescence (TL) total signal is observed to be a linear function of deposited energy. To assess the contribution to the glow curve from the high-temperature peaks, two methods were studied: ratios of peak heights (peak 7 with respect to peak 5), and ratios of areas of the deconvoluted high-temperature peaks with respect to peak 5. The ratios were evaluated as a function of dose, showing in both methods a dependence on LET and ion identity. Some of the studied ions show these ratios to be independent of dose, up to 500 mGy, while for other ions, departures from linearity up to 4.5% +/- 2.5% per 100 mGy are observed at 500 mGy. These results show that, in general, the incident radiation LET is not a parameter that can be deduced from the glow curve.