Amperometric choline biosensor fabricated through electrostatic assembly of bienzyme/polyelectrolyte hybrid layers on
Jun Wang1, Guodong Liu, Yuehe Lin
1Pacific Northwest National Laboratory, Richland, WA 99352, USA.
Abstract:
We report a flow injection amperometric choline biosensor based on the electrostatic assembly of the choline oxidase (ChO) enzyme and a bienzyme of ChO and horseradish peroxidase (HRP) onto multi-wall carbon nanotubes (MWCNT) modified glassy carbon (GC) electrodes. These choline biosensors were fabricated by immobilization of enzymes on the negatively charged MWCNT surface through alternately assembling a cationic poly(diallydimethylammonium chloride) (PDDA) layer and an enzyme layer. Using this layer-by-layer assembling approach, a bioactive nanocomposite film of PDDA/ChO/PDDA/HRP/PDDA/CNT (ChO/HRP/CNT) and PDDA/ChO/PDDA/CNT (ChO/CNT) was fabricated on the GC surface. Owing to the electrocatalytic effect of carbon nanotubes, the measurement of faradic responses resulting from enzymatic reactions has been realized at low potential with acceptable sensitivity. The ChO/HRP/CNT biosensor is more sensitive than the ChO/CNT one. Experimental parameters affecting the sensitivity of biosensors, e.g., applied potential, flow rate, etc., were optimized and potential interference was examined. The response time for this choline biosensor is fast (few seconds). The linear range of detection for the choline biosensor is from 5.0 x 10(-5) to 5.0 x 10(-3) M and the detection limit is about 1.0 x 10(-5) M.