R C Maher1, L F Cohen, J C Gallop
1The Blackett Laboratory, Imperial College London, Prince Consort Road, London SW7 2BW, United Kingdom. robert.maher@imperial.ac.uk
Abstract:
We make systematic measurements of Raman anti-Stokes/Stokes (aS/S) ratios using two different laser excitations (514 and 633 nm) of rhodamine 6G (RH6G) on dried Ag colloids over a wide range of temperatures (100 to 350 K). We show that a temperature scan allows the separation of the contributions to the aS/S ratios from resonance effects and heating/pumping, thus decoupling the two main aspects of the problem. The temperature rise is found to be larger when employing the 633 nm laser. In addition, we find evidence for mode specific vibrational pumping at higher laser power densities. We analyze our results in the framework of ongoing discussion on laser heating/pumping under surface-enhanced Raman scattering (SERS) conditions.